Study of the 16-Channel Scintillation Detector Prototype with Silicon Photomultipliers Readout
https://doi.org/10.1134/S2079562920040168
Abstract
In the Laboratory of High-Energy Physics, 16-channels prototypes of detectors (with SiPM readout produced by Ketek and Hamamatsu) were developed for their application in zero-degree calorimetry and for measuring the profile of a low-energy neutron beam. This paper presents the method for determining the voltage of the breakpoint of the noise characteristic of the silicon photomultipliers in the multichannel detector and its temperature dependence in the range from 27.7 to 39.1°C. The response of SiPM to the LED source of calibrated light pulses has been studied and the time resolution has been obtained.
Keywords
About the Authors
A. V. TishevskiyRussian Federation
141980; Moscow oblast; Dubna
Yu. V. Gurchin
Russian Federation
141980; Moscow oblast; Dubna
A. Yu. Isupov
Russian Federation
141980; Moscow oblast; Dubna
P. K. Kurilkin
Russian Federation
141980; Moscow oblast; Dubna
V. P. Ladygin
Russian Federation
141980; Moscow oblast; Dubna
S. G. Reznikov
Russian Federation
141980; Moscow oblast; Dubna
Ya. T. Skhomenko
Russian Federation
141980; Moscow oblast; Dubna
A. A. Terekhin
Russian Federation
141980; Moscow oblast; Dubna
A. N. Khrenov
Russian Federation
141980; Moscow oblast; Dubna
References
1. Timoshenko A., Finogeev M., Guber F. et al. Tech. Design Report for the MPD Experiment.
2. Ketek PM3350 Datasheet. http://www.ketek.net/products/sipm/pm3350/
3. Hamamatsu S12572-010P Datasheet. https://www.hamamatsu.com/content/dam/hamamatsu-photonics/sites/documents/99_SALES_LIBRARY/ssd/mppc_kapd9008e.pdf
4. http://hvsys.ru/images/data/news/2_small_1368802824.pdf
5. Kushpil V. et al. // J. Phys.: Conf. Ser. 2016. V. 675. P. 012039.
6. http://hvsys.ru/images/data/news/5_small_1368802948.pdf
7. Isupov A. // EPJ Web Conf. 2019. V. 204. P. 10003.
8. https://afi.jinr.ru
Review
For citations:
Tishevskiy A.V., Gurchin Yu.V., Isupov A.Yu., Kurilkin P.K., Ladygin V.P., Reznikov S.G., Skhomenko Ya.T., Terekhin A.A., Khrenov A.N. Study of the 16-Channel Scintillation Detector Prototype with Silicon Photomultipliers Readout. Nuclear Physics and Engineering. 2020;11(4):242-246. (In Russ.) https://doi.org/10.1134/S2079562920040168