Sampling-Fluctuations in Electromagnetic Calorimeters
https://doi.org/10.56304/S2079562920060275
Abstract
The paper presents the results of calculations of the sampling fluctuations in Pb–LAr electromagnetic sandwich calorimeters. The GEANT4 package was used to generate showers initiated by electrons with energies E0 from 20 to 500 GeV. It is shown that the dependence of the sampling fluctuations on E0 and the thickness of the lead absorber x can be described by the formula kxb/√E0. The exponent b weakly depends on the LAr gap width d and is close to 2/3 and k = (0.1985 ± 3) – (0.0363 ± 2)lnd[mm].